Additives & InstrumentsA member ofMeasure what you see.BYK-mac i BYK-mac i COLORManual

Strona 2

9• Align the protection cap when the instrument isnot in use.• Avoid prolonged high relative humidity and do notallow condensation water.• Do not use

Strona 3

102. System description and Delivery notesBYK-mac i COLOR measures color traditionallyat five aspecular angles (15°/25°/45°/75°/110°)with a 45° illum

Strona 4 - Table of contents

11BYK - mac i COLOR* 7032BYK - mac i COLOR Sensor 7033Comes complete with:Multi-angle spectrophotometer, black calibrationstandard, white calibration

Strona 5

123. Power supplyBefore operating the instrument for the first time,please read the operating manual and takeparticular notice of the Safety Instructi

Strona 6 - 1. Safety instructions

13Power supplyRechargeable battery packTo operate the instrument with the rechargeablebattery pack, make sure it is inserted until it locksin place.Wh

Strona 7

14Charging the rechargeable battery packThe rechargeable battery pack is charged in thedocking station. Charging time is approx. 2 hours.1. Battery pa

Strona 8

151234675Controls4. ControlsMeasurement unit1 Mode scroll wheel: switch on and menu selection2 esc button: changes between main menu and measurement

Strona 9

16Getting started5. Getting startedTurning on the unitTo turn on the unit, press the mode scroll wheel orthe operate button.The main menu appears.For

Strona 10

17The following error messages might appear duringmeasurement.A warning signal is heard and the light diodes flashred. At the same time, a message app

Strona 11

186. CalibrateCalibration InformationStandardsThe instrument comes with 2 calibration standards.The black trap is used to perform the zeroreflectance

Strona 12

1BYK-mac iBYK-mac i COLORManualPatent pending 300 000 891 E 1306BYK-Gardner GmbHLausitzer Str. 8D-82538 GeretsriedGermanyTel. 0-800-gardner(0-800-4273

Strona 13 - 3. Power supply

19CalibrateCalibration Notes• Dirt or dust in the optics area will cause aninaccurate calibration reading. Therefore, alwaysuse the protective cap, if

Strona 14

206.1 Cyan CheckChoose "Cyan Check" from the "Calibrate" menu.Open the cover of the cyan checking standard andposition the instrum

Strona 15

216.2 CalibrateTurn the unit on by pressing either the menu scrollwheel or the operate button.Choose the menu item Calibrate from the Calibratemenu.Re

Strona 16 - 4. Controls

22CalibrateWhen black trap calibration is completed, theinstrument must be placed on the white standard.Position the instrument as thoroughly on the w

Strona 17 - 5. Getting started

236.3 Effect CheckChoose "Absolute" from the "Measure" menu. Thebasic configuration is the average of threemeasurements.Open the c

Strona 18

247. Measurement techniquesIn order to obtain accurate and repeatablemeasurements, the bottom of the instrument mustbe aligned flat on the sample surf

Strona 19 - 6. Calibrate

258. MeasureSelect the menu item Measure by pressing theoperate button/scroll wheel.A selection menu appears.”Continue last series”: To continue the l

Strona 20

26Measure8.2 AbsoluteYou are asked if you like to continue the test series.Confirm by pressing the scroll wheel leads you tothe last test series.You c

Strona 21

27New StandardYou are asked to input a name. You can accept thepredefined suggestion or change the character byturning the scroll wheel. When you pres

Strona 22

28MeasureExisting standardChoose the appropriate standard from the list.You are asked to continue the last series.If you choose ”Yes”, you can directl

Strona 23

2Dear customer,thank you for having decided for a BYK-Gardnerproduct. BYK-Gardner is committed to providing youwith quality products and services. We

Strona 24

29MeasureAuto standardThe function „Auto Standard“ is displayed if at leastone standard is stored in the memory. Whenchoosing this function the closes

Strona 25 - 7. Measurement techniques

308.4 OrganizerAn organizer file defines a measurement sequencefor user guidance, e. g. measuring a car body withseveral checkzones. These files can b

Strona 26 - 8. Measure

31MeasureAfter the standard has been measured, or when it isalready stored in the instrument, the name of thecheckzone to be measured first, appears o

Strona 27

32MeasureDuring the measurement you have access toadditional functions by pressing the mode wheel.In case you took a false reading, you can delete the

Strona 28

33Delete9. DeleteSelect the menu item Delete by pressing theoperate button/scroll wheel.A selection menu appears.”Last Measurement”: To delete the las

Strona 29

349.2 CheckzoneTo delete a whole checkzone, choose the menuitem ”Delete checkzone” and press the mode wheel..The instrument is now ready again to meas

Strona 30

35DeleteTo delete test series in difference mode, you have tochoose the appropriate standard first and then thetest series to be deleted. Test series

Strona 31

3610. Data ViewYou can use this function to display the readings ofmeasurements stored in the instrument.10.1 Test seriesAbsoluteTo view the data of r

Strona 32

37Data ViewOrganizerTo view the data of an organizer, choose theappropriate organizer from the list.Then select parameter 2 (color) and the appropriat

Strona 33

38Data View10.2 StandardTo view the data of a standard, choose theappropriate standard from the menu.The data of the standard is displayed.

Strona 34 - 9. Delete

3Table of contentsTable of contents1. Safety instructions ... 52. System des

Strona 35

3911. SetupIn the Setup menu you find functions to adjustgeneral settings for:Measurement parametersInstrumentCalibrate11.1 Measurement parametersMeas

Strona 36

4011.1.1.2 SMC - Statistical Measurement ControlHighlight the function with the scroll wheel. Pressingthe wheel activates or deactivates the function

Strona 37 - 10. Data View

41Setup11.1.2.1 AnglesYou can choose individual angles or ”All”.11.1.2.2 ScalesHere you can decide according to which color scaleresults are calculate

Strona 38

42Setup11.1.2.3 Illuminant/ObserverYou can choose among the illuminants/observersshown on the left side.11.1.3 Effect (only available with BYK-mac i)T

Strona 39

43SetupTo change the character, turn the scroll wheel.When you press the wheel, the highlighted positionmoves to the next character. Pressing operatec

Strona 40 - 11. Setup

44Setup11.2 InstrumentIn the Setup menu you find functions to adjustgeneral settings of the instrument:11.2.1 Sample curvatureYou can adapt the sensit

Strona 41

45Setup11.2.3 BeeperThis menu option turns the beeper on or off. Use thescroll wheel to move the cursor to Beeper and pressthe wheel.When the beeper i

Strona 42

46Setup11.2.6 Date/TimeThe unit contains an integrated clock. This makesdate and time of the measurement available for datatransfer to a PC. Date and

Strona 43

4711.2.9 AdvancedYou can use this menu item to activate:• Interrupt statistic• Input Parameter 3• Input Comment• Input ID• Input Checkzone11.2.9.1 Int

Strona 44

48Setup11.3 CalibrateYou can perform a cyan check and a completecalibration.Please refer to the chapter ”Calibration”.

Strona 45

412. Interface...4913. Standards...

Strona 46

4912. InterfaceTransfer of data stored in the memory of theinstrument takes place through the docking station.It contains the USB interface connectin

Strona 47

5013. StandardsDIN 5033 Colorimetry; basic concepts.DIN 5036 Radiometric and photometric properties of materials;definitions characteristic.DIN 6174 C

Strona 48

5114. Technical DataGeneral technical data:Temperature range 10 °C to 42 °C (50 °F to 110 °F)for operation0 °C to +60 °C (32 °F to 140 °F )for storage

Strona 49

52Display 2.7 in. TFT color LCD displayPower supply Rechargeable battery pack or 4 mignon AA batteries(Alkaline or rechargeable)Dimensions (LxWxH) 218

Strona 50 - 12. Interface

53Errors and Warning Messages15. Errors and Warning MessagesInfoWarningThe surface The sample temperature is lower than thetemperature is under ..°C s

Strona 51 - 13. Standards

54ErrorInstrument temperature Temperature is outside the operatingout of range temperature range.Ambient light Appears when there is too much ambientl

Strona 52 - 14. Technical Data

55Cleaning and Maintenance16. Cleaning and Maintenance• Do not insert any objects into the measurementaperture for cleaning. The instrument could getd

Strona 53

56Cleaning the Black TrapFrom time to time the black trap should be cleanedwith clean, dry air to remove any dust orcontamination.The black trap is ta

Strona 54

5717. Service and CertificationServiceBesides the repair of your instrument we offer thefollowing additional services:First diagnosis on the telephon

Strona 55

58Service and CertificationService Centers for BYK-Gardner productsGermanyBYK-Gardner GmbHLausitzer Strasse 882538 GeretsriedGermanyPhone:+49-8171-34

Strona 56 - 16. Cleaning and Maintenance

5Safety instructions1. Safety instructions • Before operating the instrument the first time,please read the operating instructions and takeparticular

Strona 57

5918. CopyrightThis instruction manual is an important part of this instrument. It contains es-sential information about setting up, placing in servic

Strona 58

300 000 891 E 1306

Strona 59

6Safety instructionsThe following symbols and terms are used.This symbol warns of the danger of injury.This symbol warns of the danger of injury cause

Strona 60 - 18. Copyright

7DANGER injuries possible• Defects and extraordinary loadsIf safe operation can no longer be presumed,shut down the device and secure it againstunint

Strona 61 - 300 000 891 E 1306

8Safety instructionsb) by disconnecting the power supply unit plug fromthe socketThe power supply unit can be disconnect from thepower supply by disco