8. Continous Mode
In the normal measuring mode,
individual measurements are
started by pressing <operate>. In
Continuous Mode, on the other
hand, measurements are repeated
and updated constantly. This lets
you "run" a rather larger area of the
specimen and get an impression of
nonuniform zones.
Continuous Mode is possible only for
a single parameter.
Select
"Continuous
Mode" in the
SET-UP menu.
Select the desired parameter. The
parameter can be changed at will.
The haze measurement needs the
value for total transmittance as a
reference for the measured
specimen. When applicable, a
prompt to this effect will appear on
the display.
set up
trans-
mittance
haze clarity
Transmittance
Measurement
Please place
sample at
haze-port
Ready: OPERATE