Ref. Measurement
only in
Continuous-Mode
Memory
full
Cause
A date before 1980 was input in
SET-UP.
Solution
The input date has been ignored,
and the old date has been retained.
Input the correct date.
Cause
The <reference> button has been
pressed while the instrument is not
in Continuous Mode.
Solution
None; this message is only
informational.
Cause
The current statistics memory is full.
Solution
Use the <delete-last> or <delete-
all> button to delete the last value
or the entire memory.